As-found and as-left: the two numbers on an instrument calibration report

· 4 min read

When the calibration technician leaves, the report they hand over has two columns of readings for every instrument they touched. As-found is what the instrument read at each calibration point before anyone adjusted it. As-left is what it read afterwards, once it had been brought back into tolerance.

Most shops look at the as-left, see it is in tolerance, file the report and move on. That is the column that says the instrument is good from today. The as-found is the one that says whether it was good until today — and it is the one with consequences for work you have already shipped.

Two questions, two columns

Put plainly:

  • As-left answers: can we use this instrument now? If every as-left reading is inside the tolerance, the instrument is fit for the next interval. This is the number that later feeds anything calculated from the instrument’s accuracy — an alternate SAT, for one.
  • As-found answers: could we trust it since the last calibration? The as-found readings show how far the instrument had drifted over the interval. If they were inside tolerance, every load controlled and recorded by that instrument since the last calibration was measured by an instrument that was in tolerance. If they were not, that is no longer known.

An instrument that is found out of tolerance and adjusted back in passes its calibration. The work it controlled before the adjustment does not get the same treatment automatically.

What an out-of-tolerance as-found sets off

An as-found reading outside tolerance means the instrument was reading wrong for some part of the interval, and nobody knows which part. The responsible conclusion is that every load it governed since its last good calibration is in question until someone has looked.

That review is a specific piece of work:

  1. Find the window. From the last calibration that was found in tolerance to this one.
  2. Find the loads. Every load on that furnace, in that window, that the instrument controlled or recorded.
  3. Judge the effect. Compare the size and direction of the error against each load’s requirements. A +2 °F drift on a load with a wide band may be harmless; the same drift on a tight aging cycle may not be.
  4. Record the outcome, including any customer notifications the review decides on.

None of that is hard if the shop can answer step two quickly. In a system where loads, furnaces and instruments are linked records, it is a filtered list. In a system where the calibration report is a PDF and the load history is somewhere else, it is an afternoon with a spreadsheet — which is the same problem an auditor walking a job backwards runs into from the other direction.

Keep every point, not a pass/fail

A calibration report is tempting to summarise as one line: calibrated, passed, next due in three months. That throws away the evidence.

What should be kept, per calibration point:

  • the set point that was injected;
  • the as-found reading, and whether it was in tolerance;
  • any adjustment made;
  • the as-left reading, and whether it was in tolerance;
  • the tolerance band the readings were judged against.

The pass/fail is derived from those. Kept as data, they answer the questions that matter later: was this instrument ever found out of tolerance? How much has it drifted, calibration over calibration? Which as-left values did last week’s alternate SAT use? Kept as a summary line, they answer none of them. This is the same rule that runs through what AMS 2750 asks of software: store the measurement, derive the verdict.

Drift is a schedule signal, too

Read over several calibrations, the as-found column tells you something the as-left never will: how fast each instrument wanders. An instrument that is found close to its tolerance limit every time is not a healthy instrument with a lucky record. It is a candidate for a shorter interval, repair or replacement — before the calibration where it is found outside.

That is only visible if the points are kept from one calibration to the next.

The one-line test

Mark one instrument’s most recent calibration as found out of tolerance. Ask the system which loads that puts in question.

A system that links instruments to equipment and equipment to loads can produce the list. One that holds calibrations as a due date and a PDF cannot, and the review starts with somebody searching the schedule by hand.


Calibration intervals, accuracy requirements by instrument type and furnace class, and the formal handling of an out-of-tolerance finding come from your copy of AMS 2750, your quality system and your customers’ requirements — not this post. What does not change: as-left says the instrument is good now, as-found says whether it was good before, and only the second one reaches back into shipped work.

  • pyrometry
  • calibration
  • ams-2750
  • audits