Alternate SAT: when system accuracy is a calculation, not a probe

· 5 min read

A conventional system accuracy test is a measurement. A calibrated test thermocouple goes in beside the furnace’s own sensor, both are read, the test reading is corrected, and the difference between the two is compared with the tolerance for that furnace. If it is inside, the system — sensor, lead wire and instrument together — passes. What a SAT proves, and what it does not, is its own subject.

An alternate SAT gets to the same answer without the probe. Current revisions of AMS 2750 allow a shop to establish system accuracy by calculation: nobody inserts anything, and the result is worked out from two errors the shop has already documented. That makes it fast, and it is why shops that use it run one every week or two instead of on a long interval. It also makes it entirely dependent on the quality of the two documents it is calculated from.

The arithmetic is one addition

The system’s error is the sum of the errors of its parts:

  • The instrument’s error. From the instrument’s most recent calibration: the as-left reading at each calibration point. If 850 °F was injected and the instrument read 850.6 after adjustment, the instrument contributes +0.6 °F at 850.
  • The sensor’s error. From the calibration certificate of the wire the furnace’s sensor was cut from: the cert’s correction at that temperature. If the cert says that wire reads 1.2 °F low at 850, the sensor contributes that.

Add them at each calibration point, compare the sum with the SAT tolerance, and the system passes or it does not. There is no field work, no stabilisation wait, no reading to take.

So the arithmetic is trivial. What makes an alternate SAT trustworthy or worthless is whether those two inputs are what they claim to be.

Input one: the instrument’s latest as-left, not a remembered one

The instrument error has to come from the calibration that is actually current for that instrument, at the points that calibration actually tested. Not the calibration before last, not the nominal accuracy from the datasheet, not a number somebody carried over from last month’s alternate SAT because the form looked the same.

That sounds obvious until the alternate SAT is a form someone fills in by hand. Then the as-left values are retyped from a vendor report, and the first time the instrument is recalibrated and the form is not, every alternate SAT after it is calculated from a calibration that no longer exists. It still passes. It is just not about this instrument any more.

The fix is structural: the alternate SAT should read the instrument’s latest calibration, and the calibration should be a record with its points in it — not a PDF in a folder. Then “which calibration was this calculated from” has an answer, and recalibrating the instrument changes the input automatically. More on what those calibration numbers mean in as-found versus as-left.

Input two: the sensor’s cert, interpolated, never extrapolated

The sensor error comes from a thermocouple calibration cert, which tests the wire at a handful of temperatures — every hundred or few hundred degrees. The furnace’s calibration points rarely land exactly on them, so the correction at 850 °F has to be interpolated between the cert’s 800 and 900.

Two rules matter here, and both are easy to get wrong in a spreadsheet:

  1. Interpolate between tested points; never extrapolate past them. If the cert stops at 1600 °F and the furnace calibration point is 1750, there is no correction for 1750. The honest answer is “this cert does not cover this point”, not the 1600 value carried forward.
  2. Use the cert the sensor was actually made from. A cert covers a spool of wire, and every sensor cut from that spool inherits it. The link that matters is this sensor came from that spool — which means the shop has to know, for each furnace position, which wire is in it. We go through what the cert’s numbers are for in reading a thermocouple calibration certificate.

When to run one: the two events that change an input

Because the result is a function of two inputs, the times it can change are exactly the times either input changes:

  • A sensor is replaced. New wire, new cert, new sensor error.
  • An instrument is recalibrated. New as-left readings, new instrument error.

A system that knows about both events can say which furnaces need a fresh alternate SAT without anyone keeping a list. A system that treats alternate SATs as a calendar entry will happily show a furnace as current on an alternate SAT calculated from a sensor that was pulled out two weeks ago.

What has to survive the audit

An auditor looking at an alternate SAT is not going to redo the addition. They are going to ask where each number came from:

  • Which calibration record supplied the instrument error, and was it current on the test date?
  • Which cert supplied the sensor error, was it approved, and is that the wire in the furnace?
  • Who checked the calculation, and who signed it off?

If the alternate SAT is a document with numbers typed into it, every one of those is answered by someone’s memory. If it is built from linked records, each answer is a click — and it reproduces the same way next year, because the records it was calculated from are kept as they were. That is the same principle as the rest of the compliance model: the result is derived from evidence, and the evidence is still there.

The one-line test

Recalibrate an instrument. Ask the system which alternate SATs are now out of date.

A system that calculates alternate SATs from linked records can answer that, because it knows which calibration each one used. A system that stores them as filled-in forms cannot — and will keep showing the old result as current.


Whether and how a shop may use the alternate SAT process, the conditions attached to it, whether a periodic comparison SAT is still required, and the tolerances by furnace class all come from your copy of AMS 2750 and your customers’ requirements — not this post. The shape does not change: two documented errors, one addition, and both inputs traceable to a real record.